HD21AB , HD21AB Delta Ohm , HD21AB Delta Ohm Vietnam
Dai Ly Detal Ohm, Dai Ly Detal Ohm Vietnam, Detal Ohm Viet Nam, Detal Ohm Vietnam, HD21AB, HD21AB Delta Ohm, HD21AB Delta Ohm Vietnam,
GHM Group, GHM Group Vietnam,
Dai Ly Detal Ohm, Dai Ly Detal Ohm Vietnam, Detal Ohm, Detal Ohm Viet Nam, Detal Ohm Vietnam
[ GB ]
GHM Group, GHM Group Vietnam,
Dai Ly Detal Ohm, Dai Ly Detal Ohm Vietnam, Detal Ohm, Detal Ohm Viet Nam, Detal Ohm Vietnam
[ GB ]
HD21AB and
HD21AB17
IAQ Monitors
are bench-top/portable instruments
manufactured by Delta Ohm
for the analysis
of indoor air
quality (IAQ, Indoor Air Quality).
The instruments simultaneously measure the parameters:
• Carbon Dioxide CO2
• Carbon Monoxide CO
• Atmospheric Pressure The HD21AB17 instrument also
measures:
• Temperature
• Relative Humidity
and it calculates:
• Dew Point
• Wet Bulb Temperature
• Absolute Humidity
• Mixing Ratio
• Enthalpy
HD21AB and HD21AB17 are
dataloggers with a
memory capacity of
67600 recordings, divided in 64 blocks. They use the DeltaLog10 software
from version 0.1.5.3.
Reference Standards:
ASHRAE 62.1 – 2004,
Legislative Decree 81/2008.
These
regulations apply to
all confined spaces
that could be
used by people.
Kitchens, baths, changing rooms and swimming pools are
included, due to the
presence of high humidity. You should take into account, in
regard to air quality,
possible chemical, physical and biological contaminants.
The instruments have
a wide Dot
Matrix graphic display
with a resolution
of
160x160 dots.
The instruments typical applications are:
• Measurement of IAQ (Indoor
Air Quality) and
comfort conditions in
schools, offices and indoor spaces.
• Analysis and study of the Sick Building Syndrome, and of
the resulting consequences.
• Checking the HVAC (Heating, Ventilation
and Air Conditioning)
system efficiency.
• Examination of IAQ conditions in factories to optimize
microclimate and improve productivity.
• Building Automation checks.
Instrument Technical Data
Instrument
Dimensions
(Length x Width x Height)
210x90x40 mm (HD21AB)
300x90x40 mm (HD21AB17 with probe)
Weight
470 g (batteries included)
Materials
ABS, rubber
Display
Backlit, Dot Matrix
160x160 dots, visible area 52x42 mm
Operating conditions
Operating temperature
-5...50°C
Storage temperature
-25...65°C
Working relative humidity
0 ... 85% RH without condensation
Instrument uncertainty
± 1 digit @ 20°C
Power
Mains adapter (code SWD10)
12Vdc/1A
Batteries
4 x 1.2V Ni‑MH rechargeable batteries AA type
Autonomy
8 hours of continuous use in measure mode
Power absorbed
with instrument off
< 45μA
Security of stored data
Unlimited
Serial interface:
Socket:
mini-USB
Type:
USB 1.1 or 2.0 not insulated
Baud rate:460800
Data bits:8
Parity:None
Stop bits:1
Flow control:Xon-Xoff
Cable length:Max 5 m
Memory
Divided in 64 blocks.
Storage capacity
67600 recordings.
Logging interval
Selectable among: 15, 30 seconds, 1, 2, 5, 10,
15, 20, 30 minutes and 1 hour.
Logging interval
Storage capacity
Logging interval
Storage capacity
15 seconds
About 11 days and 17 hours
10 minutes
About 1 year and 104 days
30 seconds
About 23 days and 11 hours
15 minutes
About 1 year and 339 days
1 minute
About 46 days and 22 hours
20 minutes
About 2 years and 208 days
2 minutes
About 93 days and 21 hours
30 minutes
About 3 years and 313 days
5 minutes
About 234 days and 17 hours
1 hour
About 7 years and 261 days
Technical data of the sensors
CO
2
Carbon Dioxide
Sensor
NDIR Dual Wavelength
Measurement range
0 ... 5000ppm
Sensor working range
-5 ... 50°C
Accuracy
±50ppm+3% of measurement
Resolution
1ppm
Temperature dependence
0.1%f.s./°C
Response time (T
90
)
< 120 sec (air speed = 2m/sec)
Long-term stability
5% of measurement/5 years
HD21AB17
CO Carbon Monoxide
Sensor
Electrochemical cell
Measurement range
0 ... 500ppm
Sensor working range
-5 ... 50°C
Accuracy
±3ppm+3% of measurement
Resolution
1ppm
Response time (T
90
)
< 50 sec
Long-term stability
5% of measurement/year
Service life
> 5 years in normal environment conditions
Atmospheric Pressure Patm
Type of sensor
Piezo-resistive
Measurement range
750 ... 1100 hPa
Accuracy
±1.5 hPa @ 25°C
Resolution
1 hPa
Long-term stability
2hPa/year
Temperature drift
±3hPa with temperature -20 ... +60°C
Relative Humidity RH (HD21AB17 only)
Type of sensor
Capacitive
Sensor protection
Stainless steel grid filter (on request 20μm sin-
tered filter P6 in AISI 316 or 10μm sintered filter
P7 in PTFE)
Measurement range
0 ... 100 % RH
Sensor working range
-20 ... +60°C
Accuracy
±2% (10÷90% RH) ±2.5% in the remaining range
Resolution
0.1°C
Temperature dependence
±2% on all temperature range
Hysteresis and repeatability
1% RH
Response time (T90)
< 20 sec (air speed = 2m/sec) without filter
Long-term stability
1%/year
Temperature T (HD21AB17 only)
Type of sensor
NTC 10kΩ
Measurement range
-20 ... +60°C
Accuracy
±0.2°C ±0.15% of measurement
Resolution
0.1°C
Response time (T
90
)
< 30 sec (air speed = 2m/sec)
Long-term stability
0.1°C/year
ORDERING CODES
HD21AB:
IAQ Monitor datalogger
kit. It measures
CO, CO2 and atmospheric pressure. Equipped with:
DeltaLog10
software (version 0.1.5.3 and later)
for data download, monitor, and data processing on Personal
Computer, 4
x 1.2V NiMH rechargeable batteries, operating manual, case.
The cables
must be ordered separately.
HD21AB17:
IAQ Monitor datalogger
kit. It measures
CO, CO
2
, atmospheric
pressure,
temperature and relative
humidity. Equipped with:
DeltaLog10
software (
version 0.1.5.3 and later
) for data download, monitor, and data
processing on Personal Computer, 4 x 1.2V NiMH rechargeable
batteries,
operating manual, case.
Không có nhận xét nào:
Đăng nhận xét